Abstract
A procedure is described in which soil gas is utilized as an alternative to the 226Ra source for the supply of the radon gas required to fill a radon chamber where radon-measuring devices are calibrated. The procedure offers opportunities to vary the radon concentration within the chamber around an average value of about 500 Bq/m3, which is considered to be sufficient for calibrating indoor radon detectors. The procedure is simple and the radon source does not require radiation protection certification (for import and/or use), unlike the commercially produced standard radioactive (226Ra) sources.
| Original language | English |
|---|---|
| Pages (from-to) | 306-310 |
| Number of pages | 5 |
| Journal | Radiation Measurements |
| Volume | 44 |
| Issue number | 3 |
| DOIs | |
| State | Published - Mar 2009 |
Keywords
- Charcoal detector
- Radon
- Radon calibration chamber
- Soil gas
Funding Agency
- Kuwait Foundation for the Advancement of Sciences
Fingerprint
Dive into the research topics of 'The use of soil gas as radon source in radon chambers'. Together they form a unique fingerprint.Projects